Constant
Current: 0 to selected full scale current
Prog. Accuracy (Range): (high/med) ranges: ±0.25%
- (low) range: ±0.5%
Regulation: ±0.1% of selected full scale
Constant Resistance: Constant Resistance mode operates
in Amps/Volt, IEEE units entered in ohms or A/V
Prog. Accuracy: ±3% of selected full scale
Regulation: ±3% of selected full scale
Constant Voltage: 0 to selected selected full scale
Prog. Accuracy (Range): (high/med) ranges:±0.25%
- (low): ± 0.5%
Regulation: ±0.15% of selected full scale
Constant Power: 0 to full scale power
Prog. Accuracy: ±3% of full scale
Regulation: ±3% of full scale
ANALOG MODE
Ext. Prog: 0 to 10 Volts input yields 0 to selected full
scale loading in all
operating modes.
Input Impedance: 330k Ohms
Prog. Response: Limited by internal
adjustable slew rate limiter
PULSE MODE
Frequency: 0.06Hz to 20kHz
Accuracy: 0.1%
Duty Cycle: 10 - 90%(Analog)
Accuracy: 0.1%
|
Adjustable Slew Rate:
Max: 0 to full scale in 10µS
Min: 0 to full scale in 10mS
OUTPUT SIGNALS
Current Sample Output:
Scaling: 10 Volts = selected full scale
Accuracy: ±0.5% of selected full scale
Sync Output:
Timing: Synchronous with pulse generator.
Output: Sink with 10k pull up to +15V
PROTECTION
Current Limit:
Analog Models: Approximately 105% of selected full
scale current
Voltage Limit:
Analog Models: Load disconnect at 105% of
selected full scale voltage
Power Limit:
Analog Models: Approximately 4250 Watts
Thermal: Load disconnect at internal
temperature of 105°C
Undervoltage: Load inhibited at less than1
Volt, when enabled
MISCELLANEOUS
AC Input: User Selectable 100VAC,
120VAC, 200VAC, 240VAC,
±10%, 48 - 62 Hz @ 350W
Ambient Temp: 0°C to 40°C |
|
Like
the 2000W analog model, the RBL 800 watt analog programmable
series has no compromise
on performance, while adding a simple analog interface.
Sleek and compact, the 800W model is
ready to address all low-to-mid power load and test requirements.
The analog programmable RBL
800W series provides all modes of operation, all functions,
full scale range switching and
master/slave paralleling standard. The 800W RBL model
allows the customer the ultimate in
flexibility when it comes to decision time! Stand alone
or 19 inch rack mountable (see accessories
page 33) unit will meet or exceed all your performance,
reliability and quality expectations. |
|
OPERATING
RANGES (FULL SCALES)
Voltage: 10 Volts, 50 Volts, 100 Volts
Current: 2 Amps, 20 Amps, 120 Amps
Power: 800 Watts
Short Circuit: 0.007 Ohms max.
CONSTANT RESISTANCE RANGES
High Ohms Mode
Range 2A 20A 120A
10V 0-.1 A/V 0-1 A/V 0-6 A/V
50V 0-.02 A/V 0-.2 A/V 0-1.2 A/V
100V 0-.01 A/V 0-.1 A/V 0-.6 A/V
Low Ohms Mode
Range 2A 20A 120A
10V 0-1 A/V 0-10A/V 0-60 A/V
50V 0-.2 A/V 0-2 A/V 0-12 A/V
100V 0-.1 A/V 0-1 A/V 0-6 A/V
|
|
|
OPERATING
RANGES (FULL SCALES)
Voltage: 20 Volts, 200 Volts, 400 Volts
Current: 2 Amps, 20 Amps, 120 Amps
Power: 800 Watts
Short Circuit: 0.03 Ohms max.
CONSTANT RESISTANCE RANGES
High Ohms Mode
Range 2A 20A 120A
20V 0-.05 A/V 0-.5 A/V 0-3 A/V
200V 0-.005 A/V 0-.05 A/V 0-.3 A/V
400V 0-.0025 A/V 0-.025 A/V 0-.15 A/V
Low Ohms Mode
Range 2A 20A 120A
20V 0-.5 A/V 0-5 A/V 0-30 A/V
200V 0-.05 A/V 0-.5 A/V 0-3 A/V
400V 0-.025 A/V 0-.25 A/V 0-1.5 A/V
|
 |
|
OPERATING
RANGES (FULL SCALES)
Voltage: 20 Volts, 200 Volts, 600 Volts
Current: 2 Amps, 20 Amps, 40 Amps
Power: 800 Watts
Short Circuit: 0.035 Ohms max.
CONSTANT RESISTANCE RANGES
High Ohms Mode
Range 2A 20A 40A
20V 0-.05 A/V 0-.5 A/V 0-1 A/V
200V 0-.005 A/V 0-.05 A/V 0-.1 A/V
400V 0-.0025 A/V 0-.025 A/V 0-.05 A/V
Low Ohms Mode
Range 2A 20A 40A
20V 0-.5 A/V 0-5 A/V 0-10 A/V
200V 0-.05 A/V 0-.5 A/V 0-1 A/V
400V 0-.025 A/V 0-.25 A/V 0-.5 A/V
|
 |